Key points

  • Ask who the patients were: accuracy measured on enriched or curated image sets overstates performance, so look for a consecutive series from the setting in which you would use the tool.
  • Do not accept the area under the curve alone; look for calibration, the operating threshold, confidence intervals and results by sex, age, ethnicity and equipment.
  • Look for external validation at more than one site, ideally prospective, and expect performance to fall when the population, equipment or workflow changes.
  • Prefer studies that compare clinicians with and without the tool in real workflow and that measure patient outcomes, and watch for automation bias and deskilling.
  • Treat a tool without external validation or prospective evaluation as experimental, and expect evidence of monitoring for drift once it is in use.

Introduction

Diagnostic tools built on artificial intelligence (AI), mostly machine-learning models, now read mammograms, colonoscopy video, chest radiographs, electrocardiograms and retinal photographs in routine care, and hundreds have regulatory authorisation. Clinicians are increasingly asked to use, buy or trust these tools, usually on the strength of a published study that they have little time to appraise.

The literature on diagnostic AI grew faster than its quality. A 2020 systematic review of studies comparing deep learning with clinicians found only two published randomised trials and 81 non-randomised studies, of which only nine were prospective, most were at high risk of bias, and most made claims in the abstract that the design could not support [1]. A meta-analysis published in 2019 found that few studies tested the algorithm and the clinicians on the same external data set [2], and a living review of 232 prediction models for COVID-19 rated almost all at high risk of bias [3]. Many of these are the old problems of diagnostic accuracy research, described by Lijmer and by Whiting two decades earlier, in which case–control sampling, differential verification and partial reference standards inflate accuracy [4,5]; others arise from the way models are trained.

Since then the field has matured. Reporting guidelines specific to AI have been published for diagnostic accuracy studies, prediction models, trial protocols and reports, and early clinical evaluation, and risk-of-bias tools have been extended [6–13]. Randomised trials with patient-relevant outcomes have reported in breast screening, colonoscopy and retinal screening. What remains unclear for most clinicians is how to apply this body of guidance quickly to a single paper.

This narrative review addresses one question: which features of a published study of a diagnostic AI tool should a non-specialist clinician check to decide whether the tool is ready for their patients? It distils the evidence into six questions and a ten-item checklist that can be applied in about ten minutes (Table 1).

Methods

This is a narrative review. PubMed (MEDLINE) was searched through the National Center for Biotechnology Information E-utilities interface from inception to September 2026. Searches combined terms for artificial intelligence (artificial intelligence, machine learning, deep learning) with terms for each topic: AI-specific reporting and risk-of-bias guidelines; systematic reviews of the design and quality of diagnostic AI studies; randomised and prospective trials of AI tools in screening mammography, colonoscopy, diabetic retinopathy screening, electrocardiography, chest radiography, stroke imaging and sepsis prediction; and studies of dataset shift, data leakage, calibration, fairness, automation bias and post-deployment monitoring. English-language articles were eligible. Sources were selected by the authors for relevance to clinical appraisal, with priority given to reporting guidelines, systematic reviews and randomised trials, and the reference lists of key articles were checked. Each cited reference was verified against its PubMed record and digital object identifier.

Selection was not systematic, and no formal risk-of-bias assessment or quantitative synthesis was performed. Estimates are reported as published in the source studies, with confidence intervals where these were reported. Because no new data were analysed, no statistical software was used and missing data did not arise. A self-assessment against the Scale for the Assessment of Narrative Review Articles (SANRA) [14] is provided in Supplementary Material S1. Ethical approval was not required.

Results

Question 1: who were the patients?

Ask how the images or records were selected. Public image sets assembled for algorithm development often have labels extracted automatically from radiology reports, uncertain diagnoses and little clinical context [15], and enriched or case–control samples inflate sensitivity and specificity compared with a consecutive series of the patients who would actually be tested [4,16]. Models can also learn shortcuts: a deep learning model for detecting pneumonia on chest radiographs learned to recognise the hospital and the type of radiograph, such as portable films taken on the sickest wards, and performed worse on radiographs from other hospitals [17,18]. Spectrum, prevalence and case mix should be described as they would be for any diagnostic test [6].

Question 2: what was the reference standard, and how were the data split?

The label the model learns is only as good as the reference standard, and labels derived from radiology reports or discharge codes carry the errors of those documents. Then ask whether the test set was truly separate from the training data. Leakage, in which information from the test set influences training through duplicated patients, shared pre-processing or feature selection on the whole data set, is a widespread cause of over-optimistic results that cannot be reproduced [19,20]. Splits at the level of the patient, and temporal splits in which the model is tested on later patients, are more honest than random splits of images.

Question 3: was there external validation?

External validation remains uncommon. Of 516 studies of AI for diagnostic analysis of medical images published in 2018, only 6% validated the algorithm on external data [16]. In a systematic review of 86 deep learning algorithms for radiological diagnosis that had been tested externally, most performed worse on external data, about a quarter substantially so [21]. The fall reflects dataset shift: different scanners, protocols, populations and disease prevalence [22]. Of 130 AI devices cleared by the United States Food and Drug Administration (FDA) up to 2020, 126 had been evaluated only retrospectively and 93 did not report evaluation at more than one site [23], and other analyses of cleared devices found similar gaps, including little subgroup reporting [24,25]. External validation at more than one site, ideally prospective, is the minimum for a tool intended for wide use.

Question 4: is discrimination reported with calibration and subgroups?

The area under the receiver operating characteristic curve (AUC), the usual measure of discrimination, is not enough. Calibration, the agreement between predicted and observed risk, determines whether a threshold set in one place produces the same number of alerts in another, and it is often not reported [26]. Subgroup performance matters because models trained on unbalanced data underperform in the groups least represented: chest radiograph classifiers underdiagnosed disease more often in female, Black, Hispanic, younger and Medicaid-insured patients (Medicaid is a public insurance programme for people on low incomes in the United States) [27,28], and dermatology models performed worse on images of darker skin tones [29]. The Standards for Reporting of Diagnostic Accuracy Studies–Artificial Intelligence (STARD-AI) guideline and the Transparent Reporting of a multivariable prediction model for Individual Prognosis Or Diagnosis guideline updated for AI (TRIPOD+AI) require both to be reported [6,7].

Question 5: was the comparison realistic?

Many studies compare an algorithm with clinicians reading images without clinical information, outside their usual workflow and sometimes under time pressure that the algorithm does not share [1]. The relevant comparison is the clinician with the tool against the clinician without it, in the setting where it will be used. That comparison brings human factors into view. When radiologists were shown incorrect AI suggestions of the Breast Imaging Reporting and Data System (BI-RADS) category, their accuracy fell, and inexperienced readers were the most affected [30]. The benefit of AI assistance varies widely between radiologists and is not reliably predicted by their experience or unaided accuracy [31]. In a multicentre observational study, endoscopists who routinely used computer-aided detection had a lower adenoma detection rate in colonoscopies performed without it (22.4% versus 28.4% before exposure), one of the first reports of deskilling [32]. Multi-reader multi-case studies, prospective reader studies reported according to the Developmental and Exploratory Clinical Investigations of DEcision support systems driven by Artificial Intelligence (DECIDE-AI) guideline, and randomised trials answer this question; retrospective bench comparisons do not [10].

Question 6: was the outcome a patient outcome?

Detecting more findings is not the same as benefit. In colonoscopy, computer-aided detection (CADe) increases the adenoma detection rate in randomised trials [33,34], but the additional lesions are mostly small, the effect in pragmatic community trials and non-randomised studies has been smaller or absent [35–38], and no trial has yet shown fewer interval cancers. Guideline panels reading the same evidence therefore disagree: the BMJ living guideline makes a weak recommendation against routine use of CADe, and the American Gastroenterological Association makes no recommendation for or against it [39,40].

In sepsis, a proprietary prediction model in wide use (Epic Sepsis Model; Epic Systems Corporation, Verona, WI, USA) had, on external validation, an AUC of 0.63 (95% confidence interval [CI] 0.62 to 0.64); it missed two thirds of patients with sepsis while generating alerts for 18% of all hospitalised patients [41], and a second independent validation also found modest performance [42]. An updated version has since been validated prospectively across several centres, with better but still modest performance [43], and alert fatigue is the measurable cost of such tools [44]. By contrast, in a prospective study at five hospitals of the Targeted Real-time Early Warning System (TREWS), a machine-learning sepsis alert, patients whose alert was confirmed by a clinician within 3 hours had lower in-hospital mortality than those whose alert was not, although this design shows association rather than causation [45]. An earlier small randomised trial in two intensive care units reported lower mortality and shorter hospital stay with a machine-learning sepsis alert [46], and a cluster-randomised trial of decision support for septic shock in paediatric emergency departments reported clinical outcomes rather than alert performance [47]. Reviews find that randomised trials of AI in practice remain few, are concentrated in gastroenterology and radiology, and often under-report how clinicians interacted with the tool [48,49].

Worked example: mammography screening

The Swedish Mammography Screening with Artificial Intelligence (MASAI) trial randomised 105,915 women to AI-supported reading or to standard double reading by two radiologists. In the intervention arm, the AI system (Transpara; ScreenPoint Medical, Nijmegen, the Netherlands) triaged examinations to single or double reading and highlighted suspicious findings. AI-supported reading reduced the screen-reading workload by 44% [50], detected 29% more cancers [51], and gave a non-inferior interval cancer rate (1.55 versus 1.76 per 1000 women; proportion ratio 0.88, 95% CI 0.65 to 1.18), with higher sensitivity (80.5% versus 73.8%) and identical specificity (98.5%) [52]. A prospective paired-reader study in Stockholm [53], an analysis of how radiologists acted on examinations flagged by AI in that study [54], and a nationwide real-world implementation in Germany including more than 460,000 women [55] support these findings. Together these studies answer questions 1 to 6: consecutive screening populations, histologically confirmed cancer as the reference standard, prospective evaluation in practice, radiologists using the tool as intended, and interval cancer as the outcome. An earlier retrospective evaluation of a different system, which reported performance exceeding that of radiologists on curated data sets, could not answer them [56].

Worked example: diabetic retinopathy

Two autonomous systems have been evaluated in pivotal prospective trials in primary care against a reading-centre reference standard. The first (IDx-DR, now LumineticsCore; Digital Diagnostics, Coralville, IA, USA) had a sensitivity of 87.2% (95% CI 81.8 to 91.2) and a specificity of 90.7% (95% CI 88.3 to 92.7) and in 2018 became the first autonomous AI diagnostic system authorised by the FDA [57]; a second (EyeArt; Eyenuk, Woodland Hills, CA, USA) performed similarly in its own pivotal study [58]. A subsequent randomised trial in young people with diabetes showed that autonomous screening at the point of care increased the proportion who completed screening and follow-up [59], an outcome that matters more than the accuracy figure.

Worked example: electrocardiography

An AI algorithm that identifies low left ventricular ejection fraction from a standard 12-lead electrocardiogram increased the diagnosis of low ejection fraction in a pragmatic cluster-randomised trial in primary care in the United States [60], and increased the detection of cardiomyopathy in a pragmatic randomised trial in pregnant and postpartum women in Nigeria [61]. Both trials used the tool in clinical workflow and measured diagnoses made, not the AUC.

Worked example: chest radiography and stroke

In a randomised trial in a health screening population, AI assistance increased the detection of actionable lung nodules on chest radiographs [62], and a mixed-methods study is evaluating whether software from Qure.ai (Mumbai, India) that prioritises chest radiographs for reporting speeds the diagnosis of lung cancer [63]. Automated detection of large-vessel occlusion shortened transfer and treatment times in some stroke networks but not in others, because the gain depends on how the care team responds to the alert [64–66].

Appraisal instruments and regulation

Table 2 lists the instruments. For reporting, STARD-AI covers diagnostic accuracy studies [6], TRIPOD+AI covers prediction models [7], the Consolidated Standards of Reporting Trials–Artificial Intelligence (CONSORT-AI) extension and the Standard Protocol Items: Recommendations for Interventional Trials–Artificial Intelligence (SPIRIT-AI) extension cover trial reports and protocols [8,9], DECIDE-AI covers early clinical evaluation and human factors [10], the Checklist for Artificial Intelligence in Medical Imaging (CLAIM) covers imaging studies [13,67], the Minimum Information for Medical AI Reporting (MINIMAR) standard sets the minimum information for any medical AI report [68], and the Chatbot Assessment Reporting Tool (CHART) covers studies of chatbot health advice [69]. For risk of bias, PROBAST+AI extends the Prediction model Risk Of Bias ASsessment Tool (PROBAST) to machine-learning prediction models [11], and the Quality Assessment of Diagnostic Accuracy Studies–Artificial Intelligence (QUADAS-AI) tool extends QUADAS-2 to AI diagnostic accuracy studies, with a revised version in development [12,70].

Regulators are moving in the same direction. The European Union Artificial Intelligence Act classifies AI systems that are medical devices requiring third-party conformity assessment as high risk, with obligations for post-market monitoring [71]; the European Society of Radiology has issued consensus recommendations for surveillance after deployment [72,73]; and methods for detecting harmful data shift after deployment are being tested [74]. A tool that was accurate at approval can drift, so evidence of monitoring belongs in the appraisal of any tool that has been in use for more than a year.

Discussion

Principal findings

Most of the weaknesses in published evaluations of diagnostic AI can be exposed with six questions about patients, reference standard and data separation, external validation, calibration and subgroups, the realism of the comparison, and the outcome measured. The tools with the strongest evidence, AI-supported mammography reading, autonomous diabetic retinopathy screening and AI-enabled electrocardiography, are those evaluated prospectively, in workflow, with outcomes beyond accuracy. Tools assessed mainly on retrospective data, such as the original sepsis model, have performed worse when tested elsewhere.

Comparison with other guidance

The checklist builds on the users' guide to reading articles that use machine learning [75] and on the AI extensions of reporting and risk-of-bias instruments [6–13]. It differs in being written for the non-specialist clinician who must decide about a tool, and in placing more weight on the realism of the comparison, calibration in the local population and patient outcomes than on the technical details of model development.

Strengths and limitations

The checklist is grounded in published reporting and risk-of-bias guidelines and illustrated with randomised and prospective studies, and every reference was checked against its source record. The review has the limitations of its design. Only one database was searched, selection was not systematic and may favour well-known trials, and the field is changing quickly. Most examples come from imaging and from high-income countries, and the checklist has not itself been validated as a tool for appraisal. AI tools assisted the literature search and drafting, as declared below; the authors checked all content.

Implications for clinicians and policy

Clinicians should ask vendors and colleagues for external, ideally local, validation that reports calibration and subgroup results before relying on a tool, and should remain alert to automation bias. Health services should require prospective evaluation before purchase and continuous monitoring after deployment, and journals should require adherence to the AI reporting guidelines.

Unanswered questions

It is not yet known whether CADe reduces post-colonoscopy cancers, whether AI-supported mammography reduces breast cancer mortality, how large and how reversible deskilling is, how to monitor deployed tools for drift efficiently, and whether performance gaps between population groups can be closed. Tools based on large language models raise these questions again in a less mature evidence base [69].

Conclusion

The question for a clinician is not whether an algorithm can match an expert on a test set but whether patients seen with the tool do better than patients seen without it. Only a few tools have answered that question, most convincingly in breast screening; many in use have not. The six questions and the ten-item checklist let a reader tell the two apart from the paper alone, and the new reporting and appraisal instruments make it fair to expect authors to answer them.